Paperless recorder parameters
In order to test the isolation voltage of two adjacent thermocouple wires, we connect the ground wire of the withstand voltage machine and the ground wire of the pressure measuring device to a thermocouple together;
At the other sampling end, the positive wire of the withstand voltage machine is connected to the test end of the other thermocouple, and the positive wire of the pressure measuring device is manually measured at the connection terminal of the thermocouple.
The voltage measured in this way is the isolation voltage of the two thermocouple wires.

After all the equipment is turned on and the voltage of the withstand voltage machine is adjusted to 1200V, we can clearly monitor that the equipment is working normally, and the data collected by TP1000 through the 1704 acquisition module is stable.
The data curve interface is also stable, indicating that the high voltage has not affected the normal operation of the equipment.
Power Supply |
AC: 85V~265V, DC: 12~25V; Two power supply methods |
Power consumption |
≤25W (Actual power consumption is related to the numbers of channel) |
Numbers of channel |
1~32 channel ( Can expand to 64channels) |
Input signal |
Thermocouple: K,R,B,N,E,T,J,S,WRE5-26,WRE3-25 |
(Refer to the module selection table below: Function description) |
Thermal resistor: Pt1000, Pt100 |
DC current: ±120mA |
DC voltage: ±120V, ±20V, ±100mV ±20mV |
Isolation, withstand voltage and temperature measurement |
AC/DC 1500V (also applicable to voltage difference between channels) |
Storage interval |
1~19999s self-setup |
Recording memory |
8G(TP1000 Basic Module) |
Recording mode |
Cycle record |
Alarm type |
Upper and lower alarm, each channel has 4 types,(ultimate upper,upper,ultimate lower,lower) |
Communication |
Standard ethernet, RS485,RS232 |
(optional USB,4G/WiFi wireless communication,Standard Modbus TCP communication protocol) |
Standard Modbus RTU protocol |
Operation environment |
-20ºC~70ºC 0~90% RH (No condensation) |
Storage environment |
0ºC~50ºC 0~85%RH(No condensation) |
Host material |
Fireproof ABS |
Dimension |
288*288*200mm |
Installing opening hole size |
278*278mm+1 |
Dimension |
288*288*200mm |
Installing opening hole size |
278*278mm+1 |
Data acquisition moduel
Data sampling port characteristics |
Mechanical properties |
Number of channels: 4 differential |
Shell material: ABS |
AD level: 24-bit high-precision dedicated IC |
Shell color: gray green |
Sampling mode: Each channel is sampled independently |
Size: 130*93*42mm |
Sampling rate: 100HZ |
Housing installation method: embedded |
Channel-to-channel isolation voltage: 1500V AC/DC |
Protection level: IP40 |
Isolation voltage from power supply: 1500V DC |
Fire rating: UL94 |
Communication interface characteristics |
Certification information |
RS-485 Communication |
Passed CE, FCC, ROSH certification |
Environmental characteristics |
Working environment: -20ºC~50ºC 0~90%RH without condensation |
Storage environment: 0ºC~50ºC 0~85%RH without condensation |
Acquisition Range and Accuracy
Input Type |
Measuring range
(indication range) |
1S+0.1S |
0.01S |
DC voltage |
±120V |
-132V to+132V |
0.01% F.S.±0.002V |
0.01% F.S.±0.005V |
±20V |
-22V to+22V |
0.01% F.S.±0.002V |
0.01% F.S.±0.005V |
±100mV |
-120.0mV to+120.0mV |
0.01% F.S.±0.005mV |
0.01% F.S.±0.015mV |
±20mV |
-25mV to+25mV |
0.01% F.S.±0.005mV |
0.01% F.S.±0.015mV |
DC Current |
±120mA |
-23mA to +23mA |
0.01% F.S.±0.002mA |
0.01% F.S.±0.003mA |
Thermocouple measurement accuracy
Absolute error measurement requires standby for more than 30 minutes in a static environment with a room temperature of 25 degrees and a humidity of 20%-80%. |
Thermocouple |
K |
-60ºC to+1372ºC |
±(0.1% rdg. +1.0ºC) |
±(0.1% rdg. +1.5°C) |
J |
-200ºC to+1200ºC |
±(0.1% rdg. +1.0ºC) |
±(0.1% rdg. +1.5°C) |
E |
-200ºC to+1000ºC |
±(0.1% rdg. +1.0ºC) |
±(0.1% rdg. +1.5°C) |
T |
-200ºC to+400ºC |
±(0.1% rdg. +1.0ºC) |
±(0.1% rdg. +1.5°C) |
N |
-200ºC to+1300ºC |
±(0.1% rdg. +1.2ºC) |
±(0.1% rdg. +1.8°C) |
W |
+1500ºC to+2315ºC |
±(0.05% rdg.+1.5°C) |
±(0.05% rdg.+2.0°C) |
0ºC to+1500ºC |
±(0.05% rdg.+1.0°C) |
±(0.05% rdg. +1.5°C) |
R |
+800ºC to+1768ºC |
±(0.05% rdg.+1.5°C) |
±(0.05% rdg.+2.0°C) |
+400ºC to+800ºC |
±(0.2% rdg. +2.0°C) |
±(0.2% rdg. +2.5°C) |
S |
+800ºC to+1768ºC |
±(0.05% rdg.+1.5°C) |
±(0.05% rdg. +2.0°C) |
+400ºC to+800ºC |
±(0.2% rdg. +2.0°C) |
±(0.2% rdg. +2.5°C) |
B |
+800ºC to+1820ºC |
±(0.05% rdg.+1.5°C) |
±(0.05% rdg.+2.0°C) |
+400ºC to+800ºC |
±(0.2% rdg. +2.0°C) |
±(0.2% rdg. +2.5°C) |
Thermal resistor |
Pt100(3-wire) |
-200ºC to +660ºC |
±(0.02% rdg. +0.3°C) |
±(0.02% rdg. +0.4°C) |
Pt100(4-wire) |
-200ºC to +660ºC |
±(0.02% rdg.+0.2°C) |
±(0.02% rdg. +0.3°C) |
PT1000(3-wire) |
-200°C to +300°C |
±(0.02% rdg. +0.3°C) |
±(0.02% rdg. +0.4°C) |
PT1000(4-wire) |
-200°C to +300°C |
±(0.02% rdg. +0.2°C) |
±(0.02% rdg. +0.3°C) |
Warm-up time |
More than 30 minutes |
Thermocouple Measurements |
Including cold end compensation accuracy + away from heat source + away from wind source |
Standard operating status
<Precision above> |
Temperature 25±3ºC, humidity 50±30%RH, power supply 9-28VDC |
Environmental adaptability |
-20ºC to +50ºC, 0 to 90% RH (no condensation) |
Communication |
RS485,RJ45,CANFD,4G,ZigBee,WiFi,Thread,USB |
Product Description






Data logger testing report

Company Profile
Shenzhen Toprie Electronics Co., Ltd is a high technology enterprise engaged in manufacturing of environmental monitoring and data acquisition field.
Established in 2009, which is covering an area of over 3000 square meters, has 6 work lines and many staffs who have good research, design for techniques, quality controlling and production.
Business Scope: data logger& paperless recorder, humidity & temperature logger/sensor, gas detector, smart meter pressure sensor, water level meter, liquid level transmitter, data acquisition module (DAS& DAQ), IoT gateway, SCADA, edge computing gateway, RTU, DTU, thermal imaging camera, infrared temp sensor, electric& power& energy meter/power quality analyzer IOT solution etc.
And we have the advanced examine equipment, product technique and the perfect quality protect system. Beside our brand name products, we also have more than 16 years' ODM and OEM order experience and can accord the clients demands or requirements to personalized own brand.
Sincerity and Down-to-earth working approach' to provide clients with excellent service of processing! We sincerely welcome you to visit our company or contact us for cooperation!




FAQ
Q: Are you a factory/manufacturer?
A: Yes, We are professional factory/manufacturer of data logger, humidity & temperature logger/sensor, gas detector, pressure sensor, water level meter, liquid level transmitter, data acquisition moduel, IoT gateway, thermal imaging camera, power& energy meter/power quality analyzer etc.
Q: Can I customized my logo and different communication methods?
A: Yes,For customized logo that will be okay,but there is a small MOQ. For communication methods: GPRS, 4G, WIFI, LORA, LORAWAN etc. Please keep us informed in advance.
Q: How long is your lead time? If I want to customize instruments, how long will it take?
A: Usually, we can ship out within 3-5 days after receiving your payment. The customize device is depends on your order quantity.
Q: What payment method do you accept?
A: We accept payment method of T/T (bank transfer), Western Union, PayPal and so on.
Q: What package do you use for your products? What shipping method do you use?
A: 1. We use standard carton box for packing.
2. We ship via express courier (door to door service), by air, and by sea according to customers' requirements.
Q: How is your company's product after-sales service system?
A: The product warranty period is 12 months, we can provide lifetime maintenance. And we have professional pre-sales and after-sales engineer teams one-to-one after-sales servicethat can respond to you within 24 hours to resolve technical issues.
For more questions, please feel free to send inquiries to us.